Bain B, Forster T, Baker A
Scand J Haematol. 1983 Apr;30(4):311-6. doi: 10.1111/j.1600-0609.1983.tb01498.x.
The sensitivity of 3 bleeding time techniques to an aspirin-induced defect of platelet function has been assessed in 35 normal volunteers. A carefully standardised Ivy technique was compared with 2 commercial devices for a template bleeding time, the Simplate II and the Thrombolette. The standardised Ivy and the Simplate II were found to have a similar sensitivity; the Thrombolette tended to be less sensitive but the difference was not significant.
在35名正常志愿者中评估了3种出血时间技术对阿司匹林诱导的血小板功能缺陷的敏感性。将精心标准化的Ivy技术与2种用于模板出血时间的商用设备Simplate II和Thrombolette进行比较。发现标准化的Ivy技术和Simplate II具有相似的敏感性;Thrombolette的敏感性往往较低,但差异不显著。