Fluosol-DA, an emulsion of perfluorocarbons, is used as a blood substitute. Its presence in samples affects results of certain radioimmunoassays routinely done in our laboratory. In several, added Fluosol-DA caused an upward displacement of the standard curve, the extent of which depended on the structure and quantity of the assayed antigen, the concentration of Fluosol-DA, and the nature of the immunoreagents. Fluosol-DA markedly affected results of assays for triiodothyronine, thyroxin, and digoxin (by one method), but assays for aminoglycosides and protein hormones were unaffected. We determined that Fluosol-DA interacts with unbound labeled (and, presumably, with unlabeled) antigen to form a complex that coprecipitates with the bound antigen-antibody fraction during the separation step. In comparison with the idealized condition we observed an apparent increase in antibody-bound labeled antigen and consequently an apparent underestimation of the measured antigen. Depending on the nature of the immunoreagents, antibodies can extract more or less of their respective antigens from particles of Fluosol-DA emulsion.
氟碳代血液(Fluosol-DA)是一种全氟碳化合物乳剂,用作血液替代品。它在样本中的存在会影响我们实验室常规进行的某些放射免疫分析的结果。在几种分析中,添加氟碳代血液会导致标准曲线向上位移,其程度取决于被测抗原的结构和数量、氟碳代血液的浓度以及免疫试剂的性质。氟碳代血液显著影响三碘甲状腺原氨酸、甲状腺素和地高辛(通过一种方法)的分析结果,但对氨基糖苷类和蛋白质激素的分析结果没有影响。我们确定氟碳代血液与未结合的标记抗原(可能也与未标记抗原)相互作用形成复合物,该复合物在分离步骤中与结合的抗原-抗体部分共沉淀。与理想条件相比,我们观察到抗体结合的标记抗原明显增加,因此测得的抗原明显被低估。根据免疫试剂的性质,抗体可以或多或少地从氟碳代血液乳剂颗粒中提取其各自的抗原。