A simple microcomputer-controlled device for automated measurement of refractory period of excitable tissue.

作者信息

Bose D, Delaive J, Mazerall E W

出版信息

J Pharmacol Methods. 1982 Sep;8(2):91-8. doi: 10.1016/0160-5402(82)90062-6.

Abstract

A computer-controlled tissue stimulator interface will be described. This device can generate a basic train at a predetermined rate. An extra pulse is introduced after a selected number of pulses. The initial latency of this extra pulse is shorter than the tissue refractory period. The latency is increased, automatically, until the tissue produces a conducted response. This is made possible by the computer checking the tissue response. The device is inexpensive and can be readily controlled by any 8 bit microcomputer, such as the Commodore PET or VIC-20. The use of a signal conditioner has been discussed. Such a circuit facilitates detection of the often slowly occurring action potential of the premature beat riding on the repolarization phase of the normal action potential.

摘要

将介绍一种计算机控制的组织刺激器接口。该设备可以以预定速率生成基本脉冲序列。在选定数量的脉冲之后引入一个额外脉冲。这个额外脉冲的初始潜伏期短于组织不应期。潜伏期会自动增加,直到组织产生传导反应。这通过计算机检查组织反应得以实现。该设备价格低廉,并且可以由任何8位微型计算机轻松控制,比如Commodore PET或VIC - 20。已经讨论了信号调节器的使用。这样的电路有助于检测通常在正常动作电位复极化阶段缓慢出现的早搏动作电位。