Insertion of a multicomponent ceramic prosthesis with extracortical fixation to the conically reamed shaft of the humerus was performed in 7 patients after resection of 4 malignant and 3 benign tumors of the proximal humerus. The length of the replacement ranged from 9 to 18 cm. Fixation was secure in all but one patient, who sustained a fracture through the proximal humerus. Subluxation occurred in 3 patients but posed no difficulty, except in one patient who required revision for a dislocation. Functionally, little active motion is achieved, but the implant provides a spacer to permit efficient use of the hand and elbow. No patient has had local recurrence of the tumor. One patient died one year postoperatively from an unrelated second primary cancer. Although the preliminary results are encouraging, further investigations are necessary on the mechanical aspects of the cone fixation principle and of the techniques of shoulder muscle reconstruction to improve the function of the prosthetic joint.
7例肱骨近端4例恶性肿瘤和3例良性肿瘤切除术后患者,接受了带有皮质外固定的多组件陶瓷假体植入术,该假体植入经锥形扩髓的肱骨干。置换长度为9至18厘米。除1例患者肱骨近端发生骨折外,其余患者固定均牢固。3例患者发生半脱位,但除1例因脱位需要翻修外,其余均未造成困难。功能方面,主动活动较少,但植入物提供了一个间隔物,以便有效地使用手和肘部。没有患者出现肿瘤局部复发。1例患者术后1年死于无关的第二原发性癌症。尽管初步结果令人鼓舞,但仍有必要进一步研究锥形固定原理的力学方面以及肩部肌肉重建技术,以改善人工关节的功能。