van Straaten F J, van Aken J
Oral Surg Oral Med Oral Pathol. 1982 Sep;54(3):347-59. doi: 10.1016/0030-4220(82)90108-6.
Intraoral radiographs are often made with circular fields to irradiate the film, and in many instances these fields are much larger than the film. The feasibility of reducing a circular radiation field without increasing the probability of excessive cone cutting was evaluated clinically, and an optimum field size was determined. A circular radiation field 4.5 cm. at the tube end was found to minimize cone cutting and reduce the area of tissue irradiated by at least 44 percent. Findings suggest that current I.C.R.P. recommendations for a 6 to 7.5 cm. diameter circular field may be too liberal.