Popelka G R, Dubno J R
Acta Otolaryngol. 1978 Jul-Aug;86(1-2):64-70. doi: 10.3109/00016487809124721.
Previous studies which have measured acoustic-reflex responses to bone-conducted signals have not effectively differentiated reflex responses from artifacts. A convenient method for identifying such artifacts was developed and employed on some acoustic-reflex measures for bone-conducted signals. The findings indicated that artifacts result when a frequently-used acoustic admittance meter (Grason-Stadler 1720B) and a conventional bone vibrator were used to measure reflex responses for tonal and noise-activating signals. It was suggested that the method be employed in future studies which investigate the acoustic reflex in response to bone-conducted signals.