Mijnheer B J, Guldbakke S, Lewis V E, Broerse J J
Antoni van Leeuwenhoek Hospital, Amsterdam, The Netherlands.
Phys Med Biol. 1982 Jan;27(1):91-6. doi: 10.1088/0031-9155/27/1/009.
The sensitivity of an energy-compensated Geiger-Müller counter to neutrons with approximate energies of 0.6, 5 and 15 MeV, relative to its sensitivity to gamma rays (kU-value), has been determined at different centres in Europe. The time-of-flight, angular difference, associated-particle coincidence and lead filtration technique have been compared, using the same Geiger-Müller counter. For 15 MeV neutrons, a consistent kU-value of 1.9% was obtained. For the lower neutron energies, lower kU values were observed with reasonable agreement.