J Nucl Med. 1980 Feb;21(2):156-60.
Quantitative assessment of skin perfusion has value in both medical and surgical therapeutic decision-making. We have adapted a technique using intradermal xenon-133 dissolved in saline for use with a gamma camera interfaced to a mini-computer. It allows rapid evaluation of several sites simultaneously. The monoexponential washout rate of the injected tracer during the first 6 min is entered into the Schmidt--Kety equation to provide blood-flow rates in ml/min per 100 g. Reproducibility of the method in normal subjects is satisfactory. An important experimental variable is ambient temperature, since perfusion rates in normal limbs at low temperatures can approach those levels found in ischemic limbs. The intradermal technique has practical advantages over other methods for study of limb or skin perfusion.
皮肤灌注的定量评估在医学和外科治疗决策中均具有价值。我们采用了一种技术,即将溶解于盐水中的皮内氙 - 133与连接到微型计算机的γ相机配合使用。它能够同时对多个部位进行快速评估。将注射示踪剂在最初6分钟内的单指数洗脱率代入施密特 - 凯蒂方程,以得出每100克组织的血流速率(毫升/分钟)。该方法在正常受试者中的可重复性令人满意。一个重要的实验变量是环境温度,因为低温下正常肢体的灌注速率可接近缺血肢体的水平。皮内技术相较于其他研究肢体或皮肤灌注的方法具有实际优势。