Cook D W
J Assoc Off Anal Chem. 1981 May;64(3):771-3.
An automatic water bath incubator which can be constructed from easily obtainable materials is described. This incubator provides temperatures necessary for both the 35 +/- 0.5 degrees C resuscitation period and the 44.5 +/- 0.2 degrees C final incubation period as required in the AOAC method for enumerating fecal coliforms by the modified A-1 test. Statistical analysis of experimental data indicated no significant differences (P greater than 0.50) in fecal coliform counts obtained by using the automatic incubator and the AOAC incubation method.
本文描述了一种可用容易获取的材料制成的自动水浴培养箱。该培养箱能提供美国公职分析化学师协会(AOAC)通过改良A - 1试验对粪大肠菌群进行计数的方法中所要求的35±0.5摄氏度复苏期和44.5±0.2摄氏度最终培养期所需的温度。实验数据的统计分析表明,使用自动培养箱和AOAC培养方法获得的粪大肠菌群计数没有显著差异(P大于0.50)。