Five aneurysm clips and their respective appliers (Heifetz, Vari-Angle, McFadden, Scoville, and Yasargil) were tested for the production of small metal shards that could provoke a foreign body reaction or increase the risk of a stress-corrosion failure. Pivot and Vari-Angle-McFadden clips produced numerous large shards, the Scoville clip produced a few fine shards, and the Yasargil and Heifetz clips produced none. Metal shard production due to cold metal transfer is attributed to the abrasive mechanical interaction between clips and appliers made from metals with different degrees of hardness.

摘要

对五种动脉瘤夹及其各自的施夹器(海菲茨夹、可变角度夹、麦克法登夹、斯科维尔夹和亚萨吉尔夹)进行了测试,以检测是否会产生可能引发异物反应或增加应力腐蚀失效风险的小金属碎片。枢轴夹和可变角度 - 麦克法登夹产生了大量大碎片,斯科维尔夹产生了一些细小碎片,而亚萨吉尔夹和海菲茨夹未产生碎片。由于冷金属转移导致的金属碎片产生归因于由不同硬度金属制成的夹子和施夹器之间的研磨机械相互作用。