Strömberg L, Dalén N
Acta Orthop Scand. 1978 Oct;49(5):448-56. doi: 10.3109/17453677808993261.
The cortical atrophy induced by a rigid internal fixation plate on diaphyseal bone was studied on the femora of seven dogs. When the plate, which had been in position for 7 months without previous osteotomy, was removed, a pronounced reduction of the cortical bone was observed in the previously plated section of the diaphyseal bone. The atrophy took the form of loss of cortical bone mainly caused by endosteal resorption with enlargement of the medullary cavity. Neither periosteal resorption nor formation of woven bone under the site of the plate were observed. The process of adaption to the changed stress and strain conditions, caused by the mechanical joint, i.e., implants and bone, was studied by means of a histological technique, and was still in progress after a period of seven months.
在七只狗的股骨上研究了刚性内固定板对骨干骨的皮质萎缩情况。当该已在位7个月且之前未行截骨术的钢板被取出时,在骨干骨先前钢板固定的部位观察到皮质骨明显减少。萎缩表现为皮质骨丧失,主要由髓腔扩大导致的骨内膜吸收引起。未观察到骨膜吸收或钢板下方编织骨的形成。通过组织学技术研究了由机械连接即植入物和骨骼引起的对应力和应变条件变化的适应过程,并且在七个月的时间段后该过程仍在进行中。