Wilke T J, Turnbull P A
Ann Clin Biochem. 1982 Mar;19(Pt 2):104-7. doi: 10.1177/000456328201900208.
An evaluation of the Amerlex system for determination of total thyroxine (T4) and total triiodothyronine (T3) is described. The within- and between-batch precisions were acceptable, and analyses of quality control material and linearity studies demonstrated good accuracy at the clinical decision levels. The correlations obtained with NML and Ames T4 and T3 kit methods were highly significant. The Amerlex T4 and T3 methods are rapid, technically simple, and, coupled with excellent precision and accuracy, present significant advantages.