Reproducibility and quality assurance of two-dimensional gel electrophoresis of serum specimens.

作者信息

Tracy R P, Currie R M, Young D S

出版信息

Clin Chem. 1982 Apr;28(4 Pt 2):908-14.

PMID:7074881
Abstract

Currently we are using two different ISO-DALT two-dimensional gel electrophoresis systems, designated MC-Iso 1 and MC-Iso 2, for the analysis of serum and plasma samples. Here we report quality-assurance data for both of these systems. CV values for the slopes of the pH gradient (ISO dimension) are 5.6% of less; CV values for the slopes of the molecular-mass curves (log Mr vs relative mobility in the DALT dimension) are 3.4% or less. We examined the various steps of the analysis in detail for reproducibility and protein loss, using radiolabeled albumin, alpha 2-macroglobulin, and beta 2-microglobulin. Generally, in the first dimension, less protein enters the MC-Iso 2 gels (our routine system in which silver stain is used) than enters the MC-Iso 1 gels (our wide-range system for myeloma serum samples, in which the gel is stained with Coomassie Blue), on the average, 87% as much. The CV at this stage for both systems is 5--8%. During equilibration, considerable amounts of protein are lost (approximately 30% in 10 min) from the ISO gel, and the reproducibility is also decreased. Resolution in the DALT dimension has, in most cases, little or no effect on either recovery or reproducibility. Overall, for most proteins expected to appear in an ISO gel of a given pH range, approximately 50--60% of the starting material may be expected to reside in the sodium dodecyl sulfate slab gel, under our conditions. The two most important variables affecting recovery are the concentration of the NaOH (used as catholyte) and the pH of the starting sample. The overall CV for the process is between 8 and 12%.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验