Peters D C, Dean P N, Merrill J T
Cytometry. 1982 Mar;2(5):350-2. doi: 10.1002/cyto.990020513.
The ease of operation and versatility of a FACS II cell sorter were improved by replacing the original pulse height analyzer (PHA) with a computer-based dual parameter PHA and a single channel analyzer (SCA). An ND620 LSI-11 based PHA is now used to generate the sorting decisions. Sorting windows are created by intensifying one or more regions of any shape on the PHA single or dual parameter histograms. For three parameter operation, the third parameter SCA contols the dual parameter information sent to the analyzer. The SCA can also trigger an abort circuit during the PHA 15 microsecond analyzing dead time, improving the purity of the sorted fraction. The electronics that interfaces the new PHA to the sorter replaces all of the original FACS II analog signal processing circuitry with an interchangeable circuit board. The principles involved should apply to any cell sorter.
通过用基于计算机的双参数脉冲高度分析仪(PHA)和单通道分析仪(SCA)取代原来的脉冲高度分析仪(PHA),提高了FACS II细胞分选仪的操作简便性和通用性。现在使用基于ND620 LSI-11的PHA来做出分选决策。通过强化PHA单参数或双参数直方图上任何形状的一个或多个区域来创建分选窗口。对于三参数操作,第三个参数SCA控制发送到分析仪的双参数信息。SCA还可以在PHA 15微秒分析死区时间内触发中止电路,提高分选部分的纯度。将新的PHA与分选仪连接的电子设备用一块可互换的电路板取代了所有原来的FACS II模拟信号处理电路。所涉及的原理应适用于任何细胞分选仪。