血浆氨的酶法测定:Sigma和BMC试剂盒的评估

Enzymatic determination of plasma ammonia: evaluation of Sigma and BMC Kits.

作者信息

Pesh-Imam M, Kumar S, Willis C E

出版信息

Clin Chem. 1978 Nov;24(11):2044-6.

PMID:709845
Abstract

We compared two kit methods for the enzymatic determination of plasma ammonia. The procedures involved measuring the decrease in absorbance at 340 nm resulting from the consumption of NADH in the Sigma kit, NADPH in the BMC kit. Both kits gave acceptable results for aqueous samples, but for plasma specimens values with the Sigma kit significantly exceeded those obtained with the BMC kit. In our hands, values with the BMC kit were more accurate and precise. On the basis of these results, we conclude that the BMC kit is the better choice.

摘要

我们比较了两种用于酶法测定血浆氨的试剂盒方法。实验步骤包括测量Sigma试剂盒中因NADH消耗、BMC试剂盒中因NADPH消耗导致的340nm处吸光度的下降。两种试剂盒对水性样品均给出了可接受的结果,但对于血浆样本,Sigma试剂盒测得的值显著超过了BMC试剂盒的值。在我们的实验中,BMC试剂盒测得的值更准确、精确。基于这些结果,我们得出结论,BMC试剂盒是更好的选择。