组织等效介质中高能电子剂量分布的计算。I. 点单向源剂量函数的确定。

Calculation of high-energy electron dose distributions in tissue-equivalent media. I. Determination of the dose function of point unidirectional sources.

作者信息

Kozlov A P, Shishov V A

出版信息

Strahlentherapie. 1982 May;158(5):298-304.

PMID:7112625
Abstract

A method for the calculation of high-energy electron dose distributions in tissue-equivalent media is suggested. The total electron source on the exposed surface of the medium is regarded as a superposition of point unidirectional sources. A sufficiently simple analytical function for describing the axial and radial dose distribution produced by such point source is suggested. A method for determining this function, which make direct dose measurements with narrow beams unnecessary, is developed. The parameters of the point source function are derived from the depth dose distribution in a sufficiently broad beam and from the transverse dose profiles for rectangular fields. Equations for computing relative dose distributions in the volume of a tissue-equivalent medium with a flat surface are presented. The method is primarily intended for use in electron beam treatment planning but may find other applications as well.

摘要

提出了一种计算组织等效介质中高能电子剂量分布的方法。介质暴露表面上的总电子源被视为点单向源的叠加。提出了一个足够简单的解析函数来描述此类点源产生的轴向和径向剂量分布。开发了一种确定该函数的方法,该方法无需使用窄束进行直接剂量测量。点源函数的参数由足够宽的射束中的深度剂量分布和矩形野的横向剂量剖面得出。给出了计算具有平面的组织等效介质体积内相对剂量分布的方程。该方法主要用于电子束治疗计划,但也可能有其他应用。