Rutstein J E, Holahan J R, Lyons R M, Pope R M
J Lab Clin Med. 1978 Oct;92(4):529-35.
The latex agglutination test for FDP is widely employed clinically to aid in the diagnosis of DIC and other conditions. Of sera containing RF, 93% demonstrated positive FDP latex agglutination tests. Reducing agents in all instances destroyed the RF agglutinating capability. Futhermore, 86% of sera positive for FDP and RF became FDP-negative following reduction. Therefore RF was responsible for false-positive FDP latex agglutination tests in the majority of patients. Reduction of patient sera is a rapid, simple method to distinguish a positive FDP test from a false-positive due to RF.
FDP乳胶凝集试验在临床上被广泛用于辅助诊断弥散性血管内凝血(DIC)及其他病症。在含有类风湿因子(RF)的血清中,93%的血清FDP乳胶凝集试验呈阳性。所有情况下的还原剂都会破坏RF的凝集能力。此外,86%的FDP和RF均呈阳性的血清在还原后变为FDP阴性。因此,在大多数患者中,RF是导致FDP乳胶凝集试验出现假阳性的原因。对患者血清进行还原是一种快速、简单的方法,可将FDP阳性试验与由RF导致的假阳性区分开来。