Strain-controlled fatigue of acrylic bone cement.

作者信息

Carter D R, Gates E I, Harris W H

出版信息

J Biomed Mater Res. 1982 Sep;16(5):647-57. doi: 10.1002/jbm.820160512.

Abstract

Monotonic tensile tests and tension-compression fatigue tests were conducted of wet acrylic bone cement specimens at 37 degrees C. All testing was conducted in strain control at a strain rate of 0.02/s. Weibull analysis of the tensile tests indicated that monotonic fracture was governed more strongly by strain than stress. The number of cycles to fatigue failure was also more strongly controlled by strain amplitude than stress amplitude. Specimen porosity distribution played a major role in determining the tensile and fatigue strengths. The degree of data scatter suggests that Weibull analysis of fatigue data may be useful in developing design criteria for the surgical use of bone cement.