Vohra K G, Pradhan A S, Bhatt R C
Health Phys. 1982 Sep;43(3):391-7. doi: 10.1097/00004032-198209000-00009.
A TLD badge based on CaSO4: Dy Teflon TLD discs has been designed for personnel monitoring of radiation workers encountering X-rays of effective energies above 20 keV. The badge had two Teflon TLD discs, one of which is used bare (without any filter) and other which is used with a metal filter. By taking into account the TL readout of both of these discs, the problems of photon energy dependence as well as directional dependence of the badge have been reduced to a minimum. The uncertainty due to photon energy dependence above 29 KeV (eff.) was found to be within a value of +/- 15% (at an angle of incidence of 45 degrees). The directional dependence of the badge is within +/- 30% in the energy range from 35 KeV to 1.25 MeV. The badge is designed so as to provide approximate information on the effective photon energy of the X-ray exposures.