Talbot A
Acta Med Okayama. 1982 Oct;36(5):407-17. doi: 10.18926/AMO/30691.
This study investigated the optimal conditions for detection of nucleotides in blood using an IP-1B capillary isotachophoretic apparatus. The system used 10 mM HCl-beta-alanine (pH 4.2) as the leading electrolyte and n-caproic acid as the terminal electrolyte. Direct application of lysed red blood cells was shown to be inaccurate, and a method of deproteinization based on heat in a microwave oven was developed. The zones for 2,3-diphosphoglycerate, ATP, inorganic phosphate, and lactate were identified enzymatically by withdrawal of pure samples of each zone via a special withdrawal cell. The quantitative values obtained by isotachophoresis were also confirmed enzymatically. The technique is now available for convenient and accurate identification of these metabolites simultaneously.
本研究使用IP - 1B毛细管等速电泳仪研究了血液中核苷酸检测的最佳条件。该系统使用10 mM盐酸-β-丙氨酸(pH 4.2)作为前导电解质,正己酸作为终末电解质。结果表明,直接应用裂解的红细胞进行检测不准确,因此开发了一种基于微波炉加热的脱蛋白方法。通过特殊的进样池提取每个区域的纯样品,采用酶法鉴定2,3 - 二磷酸甘油酸、ATP、无机磷酸盐和乳酸的区域。等速电泳获得的定量值也通过酶法得到了证实。该技术现已可用于同时方便、准确地鉴定这些代谢物。