Wilkins B H, Kalra D
Arch Dis Child. 1982 Dec;57(12):948-50. doi: 10.1136/adc.57.12.948.
Blood glucose levels were estimated in 101 neonatal blood samples using three glucose test strip methods and the results compared with those from a laboratory. BM-test-glycemie 20-800 test strips and Reflotest-hypoglycemie test strips gave a rapid and reliable estimate of blood glucose level in the range 0-8 mmol/l (0-140 mg/100 ml). Dextrostix test strips tended to overestimate all blood glucose levels.