Burgess A E, Pate G
Med Phys. 1981 Jan-Feb;8(1):33-8. doi: 10.1118/1.594904.
Measurement results of the relative contribution of primary and secondary radiation are presented. Measurements were done with both monoenergetic and broad spectrum x-ray beams. A variety of phantoms were used: graphite, water, lucite, a skull phantom, and aluminum. We found that the primary radiation fraction is nearly independent of photon energy and tube potential. Calculations using a simple single Compton scattering model agree with the experiment results.