Raeside D E, Anderson D W, Galloway D C
Radiology. 1978 Dec;129(3):814-5. doi: 10.1148/129.3.814.
Thermoluminescent dosimeters were used to estimate the dose to the thyroid and the eye for routine clinical computed tomography of the brain using an EMI CT5005 scanner (20-sec data acquisition mode, bolus covering the head). The data can be summarized by the linear relations Dt = (7.5 +/- 0.8)N millirads (8 less than or equal to N less than or equal to 18) and De = (39 +/- 5)N + (1400 +/- 200)n millirads (8 less than or equal to N less than or equal to 20, 0 less than or equal to n less than or equal to 6) where Dt is the dose to the tissue anterior to the thyroid, De is the dose to the anterior tissue of the eyelid, N is the total number of CT cuts, and n is the number of cuts in which the globe of the eye is imaged.