Suppr超能文献

人工晶状体屈光力计算公式。

Intraocular lens power formulas.

作者信息

Fritz K J, Partamian L G, Leveille A S, Kiernan J P

出版信息

Ophthalmology. 1981 May;88(5):432-3. doi: 10.1016/s0161-6420(81)35010-6.

Abstract

We have developed a FORTRAN digital computer program to evaluate intraocular lens power for any thin lens and combination of corneal curvature and axial length. Partial derivatives of the power with respect to cornea power, axial length, and lens position allow estimation of error in power of the lens. These derivatives are computed along with power of the lens for six far points which may be changed with the program running. The program is written for remote access so that an ordinary telephone and portable terminal provide the ability to compute lens power.

摘要

我们开发了一个FORTRAN数字计算机程序,用于评估任何薄透镜以及角膜曲率和眼轴长度组合下的人工晶状体屈光度。屈光度相对于角膜屈光度、眼轴长度和晶状体位置的偏导数可用于估计晶状体屈光度的误差。这些偏导数与晶状体在六个远点的屈光度一起计算,这六个远点可在程序运行时更改。该程序编写为可远程访问,因此普通电话和便携式终端就能具备计算晶状体屈光度的能力。