颈椎矫形器的正确适配。

Proper fitting of the cervical orthosis.

作者信息

Fisher S V

出版信息

Arch Phys Med Rehabil. 1978 Nov;59(11):505-7.

PMID:727937
Abstract

This study evaluated the effect of fitting of the sternal occipital mandibular immobilization (SOMI) orthosis in restricting sagittal cervical spine motion. Cervical orthotic devices are usually fitted according to the subjective tolerance of the patient. Pressures were recorded after fitting in the usual manner by a certified orthotis and when fitted more loosely. Eight able-bodied adults were fitted in this manner and were then immobilized in a straight-backed chair while lateral cervical spine films were taken in neutral, flexion, and extension. Previously standardized methods were used to measure flexion-extension of the cervical spine. The resting pressures, when fitted in the usual manner, were approximately 105 mmHg at the chin and occiput. When pressure sensors were used to fit the brace loosely, the average pressure was approximately 25 mmHg. The data indicate that regardless of fit, there was no significant difference in the mobility of the entire cervical spine or at any intervertebral level. Recommendations for fitting of the SOMI are discussed.

摘要

本研究评估了胸骨枕下颌固定(SOMI)矫形器在限制颈椎矢状面活动方面的作用。颈椎矫形器械通常根据患者的主观耐受性进行适配。由专业矫形师按常规方式适配后以及更宽松地适配后记录压力。八名身体健全的成年人以这种方式适配,然后固定在直背椅上,同时拍摄颈椎中立位、前屈和后伸位的侧位片。采用先前标准化的方法测量颈椎的屈伸情况。按常规方式适配时,下巴和枕部的静息压力约为105 mmHg。当使用压力传感器更宽松地适配支具时,平均压力约为25 mmHg。数据表明,无论适配情况如何,整个颈椎或任何椎间水平的活动度均无显著差异。文中讨论了SOMI适配的相关建议。