Cervical orthoses effect on cervical spine motion: roentgenographic and goniometric method of study.

作者信息

Fisher S V, Bowar J F, Awad E A, Gullickson G

出版信息

Arch Phys Med Rehabil. 1977 Mar;58(3):109-15.

PMID:843201
Abstract

Movement of the cervical spine in the sagittal plane was studied in ten normal subjects from 20 to 30 years of age without and with four different cervical orthoses: (1) polyethylene Camp plastic collar with chin and occiput piece, (2) plastizote Philadelphia collar, (3) four-poster and (4) SOMI (sternal occipital mandibular immobilization). The effect of the orthoses on restricting sagittal motion was measured simultaneously using roentgenographic and bubble goniometric methods. The subject was immobilized in a straight back chair to eliminate trunk motion, and lateral cervical spine films were taken of each subject in neutral, flexion and extension without and with each orthotic device. Distortion forces exerted on the orthotic devices were standardized by measurement of pressures at the chin and occiput. Roentgenographic measurements of flexion and extension and anteroposterior displacement of the cervical spine were compared to the measurements obtained by bubble goniometry. The four-poster and SOMI were found to be most effective in restricting extension and flexion respectively. The polyethylene and plastizote orthoses were significantly less effective in restricting motion. The bubble goniometer is an adequate clinical tool in assessing overall flexion-extension of the cervical spine but is not so precise and does not give information on the degree of motion at an individual vertebral level.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验