Abstract

Twenty angiographic procedures involving the superior mesenteric artery (SMA) were performed with balloon occlusion catheters in dogs with chronic liver cirrhosis and portal hypertension. The following complications were found: (a) rapid balloon inflation during angiography due to a leak between the two catheter lumens (two cases); (b) failure to deflate a balloon located in the SMA (two cases); (c) balloon detachment intravascularly and during catheter insertion (one case each); and (d) rupture of the SMA (one case). Balloon detachment occurred only with re-used catheters. The safety of angiographic occlusion balloon catheters for general use is questioned. Under no circumstances should catheters be re-used.

摘要

对患有慢性肝硬化和门静脉高压的犬,使用球囊闭塞导管对肠系膜上动脉(SMA)进行了20次血管造影操作。发现了以下并发症:(a)血管造影期间由于两个导管腔之间的渗漏导致球囊快速充气(2例);(b)位于SMA的球囊无法放气(2例);(c)血管内和导管插入过程中球囊分离(各1例);以及(d)SMA破裂(1例)。球囊分离仅发生在重复使用的导管上。一般用途的血管造影闭塞球囊导管的安全性受到质疑。在任何情况下都不应重复使用导管。