Roggenkamp H G, Kiesewetter H, Spohr R, Dauer U, Busch L C
Biomed Tech (Berl). 1981 Jul-Aug;26(7-8):167-9. doi: 10.1515/bmte.1981.26.7-8.167.
For the determination of the deformability of individual red blood cells, a measuring device was developed, simulating the passage of individual red blood cells in capillaries of the nutritive microcirculatiuon. The kernel of this device, described here, is a single pore membrane 4--6 microns in diameter and 10 to 50 microns in length. The combination of a heavy ion accelerator (containing a faster shutter system) with nuclear tract technology makes possibly the manufacture of a large number of single pore membranes necessary for clinical and pharmaceutical tests.