Nicolosi V M, Morgantetti F, Desiderio C, Geraci C, Lamicela L, Gazzo G
G Batteriol Virol Immunol. 1980 Jul-Dec;73(7-12):281-93.
A new rotor for the ABAC system has been investigated that allows the main enterobacteria of urinary source to be identified (this is performed manually) and simultaneously the relative antibiogram (this is automatically carried out). The precision of this identification has been shown using control strains previously identified by three miniaturized kits: API 20 E, MICRO-ID and ENTEROPLATE. The accuracy of the ten identification tests present in the new "Identibiogramma" rotor has also been shown. Finally, the correspondence has been shown between the antibiogram performed with the automatized method and that with the KirbyBauer method. The data obtained evidence the validity of the new rotor and of the ABAC system.
已对ABAC系统的一种新型转子进行了研究,该转子能够识别尿路来源的主要肠道细菌(手动操作),同时还能自动生成相对抗菌谱。使用先前通过三种小型试剂盒(API 20 E、MICRO-ID和ENTEROPLATE)鉴定的对照菌株,已证明了这种鉴定的准确性。新型“Identibiogramma”转子中十种鉴定测试的准确性也已得到证明。最后,还展示了自动方法与KirbyBauer方法所做抗菌谱之间的一致性。所获得的数据证明了新型转子和ABAC系统的有效性。