弗莱施型呼吸流速计的线性度和频率响应。

Linearity and frequency response of Fleisch type pneumotachometers.

作者信息

Zock J P

出版信息

Pflugers Arch. 1981 Oct;391(4):345-52. doi: 10.1007/BF00581522.

DOI:10.1007/BF00581522
PMID:7312571
Abstract

Some aspects of the linearity and the frequency response of flow measurement with Fleisch type pneumotachometers are considered. The transition from a uniform velocity distribution at the entrance of the capillary tubes of the transducer to a fully developed flow profile is discussed as one of the phenomena influencing the linearity of the set-up. Besides, an expression is derived for the frequency response of the Fleisch pneumotachometer. The frequency responses as derived from measurements show good agreement with theory. In the range of the lower frequencies the frequency response of the flow measuring set-up is proved to be largely determined by the frequency response of the pressure transducer with the connecting tubes. A simple way is indicated to measure the latter. This, in combination with the calculated frequency response of the flow transducer, makes it fairly easy to determine the frequency response characteristics of a flow measuring set-up containing a Fleisch type pneumotachometer.

摘要

本文考虑了使用弗莱施型呼吸流速计进行流量测量时的线性度和频率响应的一些方面。讨论了换能器毛细管入口处从均匀速度分布到充分发展的流动剖面的转变,这是影响测量装置线性度的现象之一。此外,还推导了弗莱施呼吸流速计的频率响应表达式。测量得出的频率响应与理论结果吻合良好。在较低频率范围内,流量测量装置的频率响应在很大程度上由压力传感器及其连接管的频率响应决定。文中指出了一种测量后者的简单方法。这与流量传感器计算出的频率响应相结合,使得确定包含弗莱施型呼吸流速计的流量测量装置的频率响应特性变得相当容易。

相似文献

1
Linearity and frequency response of Fleisch type pneumotachometers.弗莱施型呼吸流速计的线性度和频率响应。
Pflugers Arch. 1981 Oct;391(4):345-52. doi: 10.1007/BF00581522.
2
Comparison of the performance of linear resistance and ultrasonic pneumotachometers at rest and during lobeline-induced hyperpnoea.线性阻力肺活量计和超声肺活量计在静息状态及洛贝林诱发的呼吸急促期间的性能比较。
Res Vet Sci. 2000 Apr;68(2):153-9. doi: 10.1053/rvsc.1999.0353.
3
BTPS calibration of heated Fleisch pneumotachometer.
Bull Eur Physiopathol Respir. 1983 Nov-Dec;19(6):635-40.
4
Improved frequency response of pneumotachometers by digital compensation.
J Appl Physiol (1985). 1990 Jan;68(1):382-6. doi: 10.1152/jappl.1990.68.1.382.
5
Correction factors for oxygen and flow-rate effects on neonatal Fleisch and Lilly pneumotachometers.新生儿Fleisch和Lilly呼吸流速计中氧气及流速效应的校正因子。
Pediatr Crit Care Med. 2003 Apr;4(2):227-32. doi: 10.1097/01.PCC.0000059334.04333.1B.
6
Improved accuracy and extended flow range for a Fleisch pneumotachograph.改进了 Fleisch 呼吸流速计的准确性并扩大了流量范围。
Med Biol Eng Comput. 1999 Jul;37(4):456-60. doi: 10.1007/BF02513330.
7
A technique for measuring frequency response of pressure, volume, and flow transducers.
J Appl Physiol Respir Environ Exerc Physiol. 1979 Aug;47(2):462-7. doi: 10.1152/jappl.1979.47.2.462.
8
Turbine flowmeter vs. Fleisch pneumotachometer: a comparative study for exercise testing.
J Appl Physiol (1985). 1987 Sep;63(3):1289-95. doi: 10.1152/jappl.1987.63.3.1289.
9
Effects of gas density on the frequency response of gas-filled pressure transducers.
J Appl Physiol Respir Environ Exerc Physiol. 1979 Sep;47(3):631-7. doi: 10.1152/jappl.1979.47.3.631.
10
Measurement of the frequency response and common-mode gain of neonatal respiratory pressure and flow measurement systems. Part 2: Results.
Clin Phys Physiol Meas. 1989 Aug;10(3):231-40. doi: 10.1088/0143-0815/10/3/003.

引用本文的文献

1
Laminar flowmeter for mechanical ventilator: Manufacturing challenge of Covid-19 pandemic.用于机械通气机的层流流量计:新冠疫情的制造挑战
Flow Meas Instrum. 2021 Dec;82:102058. doi: 10.1016/j.flowmeasinst.2021.102058. Epub 2021 Oct 1.
2
Improved accuracy and extended flow range for a Fleisch pneumotachograph.改进了 Fleisch 呼吸流速计的准确性并扩大了流量范围。
Med Biol Eng Comput. 1999 Jul;37(4):456-60. doi: 10.1007/BF02513330.

本文引用的文献

1
Method for the calculation of velocity, rate of flow and viscous drag in arteries when the pressure gradient is known.已知压力梯度时动脉中速度、血流速率和粘性阻力的计算方法。
J Physiol. 1955 Mar 28;127(3):553-63. doi: 10.1113/jphysiol.1955.sp005276.
2
The inlet length for oscillatory flow and its effects on the determination of the rate of flow in arteries.振荡流的入口长度及其对动脉血流速率测定的影响。
Phys Med Biol. 1961 Oct;6:303-17. doi: 10.1088/0031-9155/6/2/308.
3
Physiologic recording by modern instruments with particular reference to pressure recording.
现代仪器的生理记录,特别是压力记录。
Physiol Rev. 1960 Oct;40:753-88. doi: 10.1152/physrev.1960.40.4.753.
4
Evaluation of three types of respiratory flowmeters.三种呼吸流量计的评估
J Appl Physiol. 1957 Mar;10(2):210-4. doi: 10.1152/jappl.1957.10.2.210.
5
[The Pneumotachograph].[呼吸速度描记器]
Helv Physiol Pharmacol Acta. 1956;14(3):363-8.
6
Linearity and frequency response of pneumotachographs.呼吸流速计的线性度和频率响应。
J Appl Physiol. 1972 Jan;32(1):121-6. doi: 10.1152/jappl.1972.32.1.121.
7
Phase compensation of Fleisch pneumotachographs.弗莱施呼吸流速计的相位补偿
J Appl Physiol. 1974 Apr;36(4):493-5. doi: 10.1152/jappl.1974.36.4.493.
8
Heated Fleisch pneumotachometer: a calibration procedure.热式弗莱施呼吸速度计:校准程序。
J Appl Physiol. 1973 Jan;34(1):117-21. doi: 10.1152/jappl.1973.34.1.117.
9
[Frequency response of pneumotachographs].[呼吸流速仪的频率响应]
Bull Physiopathol Respir (Nancy). 1972 Nov-Dec;8(6):1363-76.
10
Errors in pneumotachography as a result of transducer design and function.由于传感器设计和功能导致的呼吸流速描记法误差。
Anesthesiology. 1973 Mar;38(3):275-9. doi: 10.1097/00000542-197303000-00014.