Mammography directed wire marking of nonpalpable breast lesions. A consecutive biopsy study of 100 patients.

作者信息

Siemssen O J, Blichert-Toft M, Pedersen M L, Olesen K P, Jensen L B, Jacobsen G K

出版信息

Acta Chir Scand. 1981;147(7):525-7.

PMID:7345843
Abstract

Among 665 patients referred for breast complaints, 100 women (15%) had mammographically detected nonpalpable lesions suspected of malignancy. Using a mammography directed hook wire marking, 115 excisional biopsies were carried out. A correct biopsy containing the lesion was obtained in 113 instances. There were 2 failures, due to displacements of the wire in one case and surgical inexperience in another case. Two patients had minor wound complications. Invasive carcinoma was found in 12 and carcinoma in situ in 4 biopsies, i.e., a cancer rate of 16% of patients and 14% of biopsies. The biopsy procedure was simple, time saving, and acceptable of patients. The excisions were small and the cosmetic results excellent. During mammography for control of wire position, and during dressing and surgical manipulations the flexible hooked wire remained as originally placed in all cases but one.

摘要

在因乳房问题前来就诊的665例患者中,100名女性(15%)乳房X线检查发现有可疑恶性的触诊不到的病变。采用乳房X线引导下钩丝定位,进行了115次切除活检。113例成功获取了包含病变的活检标本。有2例失败,1例是由于导丝移位,另1例是由于手术经验不足。2例患者出现轻微伤口并发症。12例活检发现浸润性癌,4例发现原位癌,即患者的癌症检出率为16%,活检标本的癌症检出率为14%。活检操作简单、省时,患者易于接受。切除范围小,美容效果极佳。在乳房X线检查以确认导丝位置期间,以及换药和手术操作期间,除1例之外,所有病例中柔性钩丝均保持原位。