Absolute induction angiometer.

作者信息

Kolin A

出版信息

Blood Vessels. 1980;17(2):61-77. doi: 10.1159/000158236.

Abstract

The objective of this method is the continuous recording of vascular diameters and their measurement in absolute terms. This can be accomplished by percutaneously introduced intravascular sensors (IVS) or surgically implanted perivascular sensors (PVS). Both types of sensor are made of bifilar insulated wires forming a bifilar loop. One wire of the bifilar loop acts as a transformer primary inducing an electromotive force (EMF) in the secondary loop wire. The latter EMF is a measure of the vascular diameter. The IVS loop is made of resilient wire and is disposed across a vascular diameter in the central plane of a blood vessel. The sensitivity is such that phasic variations of a few micrometers in the diameter of a major blood vessel (such as the aorta or a coronary artery) can be measured and recorded. The surgically implanted perivascular loop sensor permits observation of vasomotion in a conscious, unrestrained animal. Both IVS and PVS permit observation of coronary vasomotion in the beating heart. Unlike the relative induction angiometer, the present device does not require an extracorporeal magnet.

摘要

该方法的目的是连续记录血管直径并以绝对值进行测量。这可以通过经皮引入的血管内传感器(IVS)或手术植入的血管周围传感器(PVS)来实现。这两种类型的传感器均由形成双线环的双线绝缘电线制成。双线环中的一根电线充当变压器初级线圈,在次级环线中感应出电动势(EMF)。后一种电动势是血管直径的一种度量。IVS环由弹性电线制成,并横跨血管中心平面内的血管直径布置。其灵敏度足以测量和记录主要血管(如主动脉或冠状动脉)直径几微米的相位变化。手术植入的血管周围环传感器可在清醒、不受约束的动物中观察血管运动。IVS和PVS都可观察跳动心脏中的冠状动脉运动。与相对感应血管计不同,本装置不需要体外磁体。