Ackmann J J, Antonich F J, Halbach R E
Med Instrum. 1980 May-Jun;14(3):178-80.
A stimulator for evoked potential recording can become synchronized with the fundamental or harmonics of the line frequency, thus introducing interference which cannot be removed by averaging. Synchronizing the stimulus alternately to the positive-going and negative-going zero-crossing of the line will reduce 60 Hz and odd harmonics, but will not reduce higher even-harmonics. This communication describes a circuit that synchronizes the stimulus to the alternate phases of the line, but also introduces a random delay between zero and one-half the line period, thus reducing the fundamental and higher harmonics as well.
用于诱发电位记录的刺激器可能会与电源频率的基波或谐波同步,从而引入无法通过平均消除的干扰。将刺激交替同步到电源的正负过零点将减少60Hz和奇次谐波,但不会减少更高的偶次谐波。本通信描述了一种将刺激与电源的交替相位同步的电路,但也会在零到电源周期的一半之间引入随机延迟,从而减少基波和更高的谐波。