Ackmann J J, Elko P P, Wu S J
Int J Biomed Comput. 1979 Aug;10(4):291-303. doi: 10.1016/0020-7101(79)90002-3.
Evoked potentials recorded automatically at frequent intervals are a useful adjunct for monitoring head injury patients; however, unaveraged residual noise due to patient movement and synchronization of th- stimulus to harmonics of the line frequency is sometimes present. A frequency analysis was performed on 23 records with varying degrees of unaveraged noise and the results were used to design a digital filter. The frequency content of the records analyzed was largely contained in a band from D.C. to 200 Hz. A zero phase shift lowpass digital filter with a cutoff frequency of 200 Hz eliminated the majority of noise. Despite the low frequency content, a wide-band amplifier is still required to avoid waveform distortion, particularly latencies. Digital filters can then be applied without causing distortion.
频繁自动记录的诱发电位是监测头部受伤患者的一种有用辅助手段;然而,有时会出现因患者移动以及刺激与电源频率谐波同步而产生的未平均残余噪声。对23份具有不同程度未平均噪声的记录进行了频率分析,结果用于设计一种数字滤波器。所分析记录的频率成分大多包含在从直流到200赫兹的频段内。截止频率为200赫兹的零相位偏移低通数字滤波器消除了大部分噪声。尽管频率成分较低,但仍需要宽带放大器以避免波形失真,尤其是潜伏期失真。然后可以应用数字滤波器而不会导致失真。