Radojković S, Kamenica S, Jasović M, Draganić M
Cardiovasc Intervent Radiol. 1980;3(3):153-5. doi: 10.1007/BF02551978.
In one case a steel coil similar to the Gianturco coil accidentally lodged in the outflow tract of the right ventricle during the embolization of a pulmonary arteriovenous fistula. Extraction of the coil with an intravascular foreign body retrieval set (Cook Inc., Bloomington Ind.) was unsuccessful because the extraction wire broke. An improvised device was then employed, using a Cordis 8 French multipurpose catheter and a single stainless steel wire designed for surgical osteosynthesis. The steel coil was successfully caught by the device and pulled out via the femoral vein.
在一个病例中,在对肺动静脉瘘进行栓塞时,一个类似于Gianturco弹簧圈的钢圈意外地卡在了右心室流出道。使用血管内异物取出装置(库克公司,印第安纳州布卢明顿)取出该弹簧圈未成功,因为取出钢丝断裂。随后采用了一个简易装置,使用Cordis 8F多用途导管和一根专为外科骨固定设计的不锈钢丝。该钢圈被该装置成功捕获并通过股静脉拉出。