Scand J Clin Lab Invest. 1980 Jun;40(4):305-9. doi: 10.3109/00365518009092648.
When performing the ALAT-determination as recommended by the Scandinavian Committee on Enzymes, turbidity will develop due to a low ionic strength of the buffer. At the end of the pre-incubation period, the turbidity may still be developing, increasing the absorbance and lowering the ALAT-activity by 0-5 U/l. Turbidity and the error introduced in the assay, can be eliminated by increasing the ionic strength of the buffer by addition of NaCl.