Suppr超能文献

五种类型阻抗仪同侧听觉反射刺激的真耳校准

Real-ear calibration of ipsilateral acoustic reflex stimuli from five types of impedance meter.

作者信息

Lutman M E

出版信息

Scand Audiol. 1980;9(3):137-45. doi: 10.3109/01050398009076347.

Abstract

The Hearing Level scale has been established in national and international standards for tones presented via supra-aural earphones. However, the standards do not encompass stimuli from the probes of ipsilateral impedance meters. This study transfers the calibration of standardised earphones to the probes of five types of impedance meter and therefore enables the intensities of their ipsilateral stimuli to be defined in terms of Hearing Levels. The method used for transfer of calibration is based on equality of acoustic reflex responses elicitied by the impedance meter probe and by a standardised earphone. Ten subjects were tested with all five instruments to ensure comparability between instruments. The resultant calibration data are expressed in terms of Reference Equivalent Threshold Sound Pressure Levels measured in both 2 cc and Zwislocki couplers. The results indicate a considerable variation amongst instruments emphasising the need for a set of calibration data for each design of probe. This variation is less marked for the Zwislock coupler than for the 2 cc coupler.

摘要

听力级量表已在国家和国际标准中确立,用于通过 supra-aural 耳机呈现的纯音。然而,这些标准并未涵盖来自同侧阻抗计探头的刺激。本研究将标准化耳机的校准转换至五种类型阻抗计的探头,从而能够根据听力级来定义其同侧刺激的强度。用于校准转换的方法基于阻抗计探头和标准化耳机引发的声反射反应的等效性。使用所有五种仪器对十名受试者进行测试,以确保仪器之间的可比性。所得校准数据以在 2 cc 和 Zwislocki 耦合器中测量的参考等效阈值声压级表示。结果表明仪器之间存在相当大的差异,这凸显了为每种探头设计提供一组校准数据的必要性。对于 Zwislock 耦合器,这种差异比 2 cc 耦合器要小。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验