Aubry-Frize M, Leduc A
Med Instrum. 1980 Sep-Oct;14(5):272-5.
A model was developed to quantitatively predict skin and tissue temperature elevations as a function of the amplitude and duration of the electrosurgical current applied, the geometry of the electrodes used, and the skin and tissue properties of the subjects. The temperature measurements were performed with thermographic equipment. The experimental data fit the model, thus fewer measurements will now be required for studies of other electrode designs.
开发了一种模型,用于根据施加的电外科电流的幅度和持续时间、所用电极的几何形状以及受试者的皮肤和组织特性,定量预测皮肤和组织温度升高。使用热成像设备进行温度测量。实验数据与该模型拟合,因此对于其他电极设计的研究,现在所需的测量次数将减少。