使用锝-99m-甲氧基异丁基异腈/氟-18-氟代脱氧葡萄糖通过同时双同位素采集单光子发射计算机断层扫描进行静息心肌灌注/代谢成像。

Rest myocardial perfusion/metabolism imaging using simultaneous dual-isotope acquisition SPECT with technetium-99m-MIBI/fluorine-18-FDG.

作者信息

Delbeke D, Videlefsky S, Patton J A, Campbell M G, Martin W H, Ohana I, Sandler M P

机构信息

Department of Radiology and Radiological Sciences, Vanderbilt University Medical Center, Nashville, Tennessee 37232-2675, USA.

出版信息

J Nucl Med. 1995 Nov;36(11):2110-9.

PMID:7472607
Abstract

UNLABELLED

The purpose of this study was to develop a dual-isotope simultaneous acquisition (DISA) protocol using a multihead SPECT camera equipped with an ultrahigh-energy (UHE) collimator to evaluate simultaneously rest cardiac perfusion and metabolism with 99mTc-MIBI/18FDG.

METHODS

Physical measurements were first performed with phantoms to develop the acquisition protocol. Fifteen patients underwent DISA-SPECT with 99mTc-MIBI/18FDG to validate the protocol. To evaluate the quality of the 99mTc-MIBI images acquired with the UHE collimator, four patients underwent a resting 99mTc-MIBI scan acquired with a high-resolution, low-energy collimator prior to DISA-SPECT.

RESULTS

With a window of 20% for both photopeaks and a 99mTc/18F concentration ratio of 3.2:1, the spillover from 18F into the 99mTc window is 6% of the counts in the window for normal subjects. Phantom images clearly demonstrated defects measuring 2 x 1 and 2 x 0.5 cm. Technetium-99m-MIBI images obtained with the UHE and high-resolution collimators provided similar diagnostic information. Using a stenosis of > 70% as criteria to diagnose coronary artery disease, DISA-SPECT had a sensitivity of 100% and a positive predictive value of 93%.

CONCLUSION

Simultaneous evaluation of rest myocardial perfusion/metabolism with a multihead SPECT camera equipped with an UHE collimator is possible using 99mTc-MIBI/18FDG with a dual-isotope simultaneous acquisition protocol.

摘要

未标注

本研究的目的是开发一种双同位素同时采集(DISA)协议,使用配备超高能(UHE)准直器的多头SPECT相机,用99mTc-MIBI/18FDG同时评估静息心肌灌注和代谢。

方法

首先用模体进行物理测量以制定采集协议。15例患者接受了99mTc-MIBI/18FDG的DISA-SPECT检查以验证该协议。为了评估用UHE准直器采集的99mTc-MIBI图像的质量,4例患者在DISA-SPECT之前接受了用高分辨率、低能准直器采集的静息99mTc-MIBI扫描。

结果

对于两个光电峰均采用20%的窗宽,99mTc/18F浓度比为3.2:1时,正常受试者中18F溢出到99mTc窗内的计数占窗内计数的6%。模体图像清楚地显示出尺寸为2×1和2×0.5 cm的缺损。用UHE和高分辨率准直器获得的99mTc-MIBI图像提供了相似的诊断信息。以>70%的狭窄作为诊断冠状动脉疾病的标准,DISA-SPECT的灵敏度为100%,阳性预测值为93%。

结论

使用配备UHE准直器的多头SPECT相机,采用双同位素同时采集协议,用99mTc-MIBI/18FDG同时评估静息心肌灌注/代谢是可行的。