Artifacts in functional magnetic resonance imaging from gaseous oxygen.

作者信息

Bates S, Yetkin Z, Jesmanowicz A, Hyde J S, Bandettini P A, Estkowski L, Haughton V M

机构信息

Department of Radiology, Medical College of Wisconsin, Milwaukee 53226, USA.

出版信息

J Magn Reson Imaging. 1995 Jul-Aug;5(4):443-5. doi: 10.1002/jmri.1880050413.

Abstract

Unexpectedly large fluctuations in signal intensity were identified in the functional MRI (FMRI) of normal subjects breathing pure oxygen intermittently. To test the hypothesis that the signal changes were due to fluctuating concentrations of gaseous (paramagnetic) oxygen in the magnetic field, echo planar gradient echo images were acquired of a phantom contiguous to an oxygen mask through which pure oxygen was administered intermittently via plastic tubing. As a control, room air was administered intermittently or oxygen continuously in the same experimental protocol. Signal intensity changes of up to 60% temporally correlated with the administration of oxygen were produced in the phantom. In functional images prepared from the echo planar images, the signal intensity changes resulted in artifacts especially at interfaces in the phantom. The intermittent administration of pure oxygen during acquisition of data for FMRI may produce signal intensity changes that stimulate or obscure function.

摘要

在间歇性呼吸纯氧的正常受试者的功能磁共振成像(FMRI)中,发现信号强度出现了意外的大幅波动。为了验证信号变化是由于磁场中气态(顺磁性)氧浓度波动所致的假设,通过塑料管间歇性地输送纯氧,利用与氧气面罩相邻的模型采集了回波平面梯度回波图像。作为对照,在相同的实验方案中,间歇性地输送室内空气或持续输送氧气。模型中产生了高达60%的信号强度变化,且在时间上与氧气的输送相关。在从回波平面图像制备的功能图像中,信号强度变化产生了伪影,尤其是在模型的界面处。在FMRI数据采集过程中间歇性地输送纯氧可能会产生刺激或掩盖功能的信号强度变化。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验