[通过外部刺激使植入式按需起搏器失活]

[Inactivation of implanted demand pacemaker through external stimulation].

作者信息

Fill H, Gmeiner R

出版信息

Acta Med Austriaca. 1978;5(4-5):134-6.

PMID:756122
Abstract

External chest wall stimulation was performed in 87 patients in order to overdrive and thereby suppress the implanted pacemaker. The impossibility to inactivate the implanted pacer suggested a disturbed "sensing" function in 12 patients. Marked changes of the escape rhythm were found in 77 patients with suppressable pacemaker when compared with the tracings before pacemaker implantation. Increasing degree of A-V block or marked changes in heart rate were observed in 14 of 35 cases with advanced A-V block. Minor alterations were noted in cases with sick sinus syndrome and atrial fibrillation. Overdrive suppression is a helpful method for the selection of pacemaker dependent patients.

摘要

对87例患者进行胸壁外刺激,以超速驱动从而抑制植入的起搏器。12例患者无法使植入的起搏器失活,提示存在“感知”功能障碍。与起搏器植入前的心电图相比,77例具有可抑制起搏器的患者出现了逸搏节律的明显变化。在35例重度房室传导阻滞患者中,14例观察到房室传导阻滞程度加重或心率明显变化。病态窦房结综合征和心房颤动患者有轻微改变。超速抑制是选择起搏器依赖患者的一种有用方法。