用于全血葡萄糖检测的APEC分析仪评估。在儿科环境中的测试。

Evaluation of the APEC analyzer for whole blood glucose. Testing in a pediatric setting.

作者信息

Harris N S, Chmil M E, Law T, Rifai N

机构信息

Department of Laboratory Medicine, Children's Hospital, Boston, Massachusetts 02115, USA.

出版信息

Am J Clin Pathol. 1995 Oct;104(4):477-9. doi: 10.1093/ajcp/104.4.477.

Abstract

The APEC glucose analyzer (APEC, Danvers, MA) was analyzed in a pediatric clinical chemistry laboratory and its performance was compared with glucose determinations on the YSI 2300 STAT analyzer (Yellow Springs Instrument [YSI], Yellow Springs, OH). One hundred sixty-nine samples were simultaneously analyzed by the APEC and the YSI systems. The agreement between the two analyzers was excellent, with r = 0.986, slope = 1.04, and y intercept = -0.094. Analysis of the run-to-run precision yielded a CV of < or = 4.1% for both the APEC and the YSI instruments. Both analyzers displayed a sensitivity to at least 1.1 mmol/L (20 mg/dL) and both produced a linear response up to 27.8 mmol/L (500 mg/dL). Neither instrument was affected by icteric or hemolyzed samples. Variations in the patient's hematocrit did not lower the whole blood glucose values when these were compared with the equivalent plasma levels. In conclusion, the APEC glucose analyzer is suited for whole blood glucose testing in pediatric settings.

摘要

在一家儿科临床化学实验室对APEC血糖仪(APEC,丹弗斯,马萨诸塞州)进行了分析,并将其性能与YSI 2300 STAT血糖仪(黄泉仪器公司[YSI],黄泉,俄亥俄州)的葡萄糖测定结果进行了比较。APEC和YSI系统同时对169份样本进行了分析。两台分析仪之间的一致性非常好,r = 0.986,斜率 = 1.04,截距 = -0.094。对两台仪器的批间精密度分析得出,APEC和YSI仪器的变异系数均≤4.1%。两台分析仪对至少1.1 mmol/L(20 mg/dL)的葡萄糖具有敏感性,并且在高达27.8 mmol/L(500 mg/dL)的范围内均产生线性响应。两台仪器均不受黄疸或溶血样本的影响。与等效血浆水平相比,患者血细胞比容的变化并未降低全血葡萄糖值。总之,APEC血糖仪适用于儿科环境中的全血葡萄糖检测。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验