Fotuhi P C, Ideker R E, Idriss S F, Callihan R L, Walker R G, Alt E U
Medical Clinic I, Charité Hospital, Berlin, Germany.
Circulation. 1995 Nov 15;92(10):3082-8. doi: 10.1161/01.cir.92.10.3082.
In previous studies, epicardial patch electrodes decreased transthoracic defibrillation efficacy. We studied the effects of two inactive epicardial 14-cm2 titanium mesh patches on defibrillation energy requirements with nonthoracotomy internal lead configurations.
A 6/6-millisecond biphasic shock wave-form was delivered via several electrode configurations 10 seconds after ventricular fibrillation was initiated with a 60-Hz generator. In two series, a total of 16 dogs (weight, 23.3 +/- 2.4 kg) underwent an up-down defibrillation protocol. In the first series, the defibrillation threshold (DFT) was determined for each electrode configuration in the presence of two inactive epicardial patches. In the second series, DFTs were determined in the presence of an inactive right ventricular (RV) or left ventricular (LV) patch alone. For several nonthoracotomy lead configurations tested in the first 8 dogs, the mean +/- SD DFT energy increased 49% to 97% with two inactive patches on the heart compared with no patches on the heart as follows: RV to superior vena caval (SVC) electrode, from 8.9 +/- 2.6 to 18.0 +/- 14.3 J; RV to SVC plus subcutaneous array electrode, from 7.0 +/- 2.4 to 10.7 +/- 5.3 J; RV to subcutaneous pectoral plate electrode, from 6.2 +/- 1.3 to 11.4 +/- 4.0 J (P < or = .05). The lowest DFT was achieved by defibrillating between the epicardial patches (3.8 +/- 3.3 J). The second series showed that DFT voltage requirements increased significantly for all three nonthoracotomy lead configurations with the inactive LV patch alone (P < or = .05) but not with the inactive RV patch alone.
Inactive epicardial patches can significantly increase the defibrillation energy requirements for nonthoracotomy lead configurations. This negative impact may be due to an insulating effect of the patches and to a disturbance of the potential gradient field under the patches. If the same holds true in patients, these results have clinical implications. Functioning epicardial patch leads should be incorporated in the defibrillation lead system if already present. If the LV patch is nonfunctioning, such as because of a lead fracture, the marked increase in DFT due to an inactive LV patch calls for thorough DFT testing during surgery and, in selected patients, may necessitate patch removal to produce an effective transvenous-based system.
在以往的研究中,心外膜补片电极降低了经胸除颤的疗效。我们研究了两个无活性的14平方厘米钛网心外膜补片对采用非开胸体内导联配置时除颤能量需求的影响。
在用60赫兹发生器诱发室颤10秒后,通过几种电极配置发放6/6毫秒双相冲击波形式的电击。在两个系列中,共有16只狗(体重23.3±2.4千克)接受了上下除颤方案。在第一个系列中,在存在两个无活性心外膜补片的情况下,为每种电极配置测定除颤阈值(DFT)。在第二个系列中,分别在单独存在无活性右心室(RV)或左心室(LV)补片的情况下测定DFT。对于在前8只狗中测试的几种非开胸导联配置,与心脏上无补片相比,心脏上有两个无活性补片时,平均±标准差DFT能量增加了49%至97%,如下所示:右心室至 superior vena caval(SVC)电极,从8.9±2.6焦耳增至18.0±14.3焦耳;右心室至SVC加皮下阵列电极,从7.0±2.4焦耳增至10.7±5.3焦耳;右心室至皮下胸壁板电极,从6.2±1.3焦耳增至11.4±4.0焦耳(P≤0.05)。在心外膜补片之间进行除颤可实现最低DFT(3.8±3.3焦耳)。第二个系列表明,对于所有三种非开胸导联配置,单独使用无活性LV补片时DFT电压需求显著增加(P≤0.05),而单独使用无活性RV补片时则不然。
无活性心外膜补片可显著增加非开胸导联配置的除颤能量需求。这种负面影响可能是由于补片的绝缘作用以及补片下方电位梯度场的干扰。如果在患者中情况相同,这些结果具有临床意义。如果已经存在,应将正常工作的心外膜补片导联纳入除颤导联系统。如果LV补片不起作用,例如由于导联断裂,无活性LV补片导致的DFT显著增加要求在手术期间进行全面的DFT测试,并且在选定的患者中,可能需要移除补片以建立有效的经静脉系统。