Berman S S, Hunter G C, Smyth S H, Erdoes L S, McIntyre K E, Bernhard V M
Section of Vascular Surgery, University of Arizona Health Sciences Center, Tucson, USA.
Surgery. 1995 Jul;118(1):8-15. doi: 10.1016/s0039-6060(05)80003-0.
This study was undertaken to assess the application of computed tomography (CT) for surveillance of aortic grafts.
Demographics, operative technique, and graft type and size at the time of implantation of aortic grafts in 178 patients were recorded. CT measurements of graft diameters were made with calipers. Data were analyzed by analysis of variance, multiple regression, and chi-squared methods.
One hundred twenty-eight (72%) bifurcated grafts and 50 (28%) tube grafts were placed for aneurysmal disease (49%), aortoiliac occlusive disease (47%), ruptured aneurysm (2.3%), anastomotic aneurysm (1%), and graft aneurysm (0.6%). Mean implant time was 43.3 +/- 3.2 months. A total of 143 Dacron prostheses (74 woven, 69 knitted) and 35 polytetrafluoroethylene prostheses were placed. Mean percentage dilation was 49.2 +/- 4.0 for knitted prostheses, 28.5 +/- 3.0 for woven prostheses, and 20.6 +/- 1.9 for polytetrafluoroethylene prostheses compared with the graft implant size. A significant correlation was seen between graft dilation (more than 50%) and graft construction with knitted prostheses (p < 0.01, Tukey's range test). Complications detected by CT occurred in 24 (13.5%) patients including supragraft aneurysms (seven), distal anastomotic aneurysms (five), proximal anastomotic aneurysms (three), graft infections (two), perigraft fluid collections (two), graft aneurysm with thrombus and distal embolization (two), and nonvascular complications (three).
CT is a useful modality for postoperative imaging of aortic prostheses. Routine surveillance may detect complications before they become clinically apparent.
本研究旨在评估计算机断层扫描(CT)在主动脉移植物监测中的应用。
记录178例患者主动脉移植物植入时的人口统计学资料、手术技术以及移植物类型和尺寸。用卡尺对移植物直径进行CT测量。采用方差分析、多元回归和卡方检验方法对数据进行分析。
为治疗动脉瘤疾病(49%)、主髂动脉闭塞性疾病(47%)、破裂动脉瘤(2.3%)、吻合口动脉瘤(1%)和移植物动脉瘤(0.6%),共植入128个(72%)分叉移植物和50个(28%)直管移植物。平均植入时间为43.3±3.2个月。共植入143个涤纶人工血管(74个编织型、69个针织型)和35个聚四氟乙烯人工血管。与移植物植入时的尺寸相比,针织人工血管的平均扩张百分比为49.2±4.0,编织人工血管为28.5±3.0,聚四氟乙烯人工血管为20.6±1.9。移植物扩张(超过50%)与针织人工血管的结构之间存在显著相关性(p<0.01,Tukey范围检验)。CT检测到的并发症发生在24例(13.5%)患者中,包括移植物上方动脉瘤(7例)远端吻合口动脉瘤(5例)、近端吻合口动脉瘤(3例)、移植物感染(2例)、移植物周围积液(2例)、伴有血栓和远端栓塞的移植物动脉瘤(2例)以及非血管并发症(3例)。
CT是主动脉人工血管术后成像的一种有用方式。常规监测可在并发症出现临床症状之前检测到它们。