小血管中空间分辨率有限的磁共振血容量流量测量的体内验证。

In vivo validation of magnetic resonance blood volume flow measurements with limited spatial resolution in small vessels.

作者信息

Hofman M B, Visser F C, van Rossum A C, Vink Q M, Sprenger M, Westerhof N

机构信息

Department of Physiology, Free University, Amsterdam, The Netherlands.

出版信息

Magn Reson Med. 1995 Jun;33(6):778-84. doi: 10.1002/mrm.1910330606.

Abstract

The accuracy of magnetic resonance phase contrast volume flow measurements in small blood vessels is expected to be smaller than in large vessels, because of partial volume effects at the vessel boundary. Accuracy was validated in the dog femoral artery, diameter 3.5 +/- 0.7 mm, using an ultrasonic transit-time flowmeter (TT). The number of pixels per vessel diameter (ND) ranged from 1.6 to 4.8. The vessel cross-section was determined using a threshold in the magnitude image. Between the two methods the correlation coefficient was 0.95 (range 10-200 ml/min). The proportional difference (PD), (QTT-QMR)/1/2(QTT+QMR), was 0.8%, showing no systematic difference between the methods. The PDs standard deviation was 27%, and 19% for flow rates above 30 ml/min. Only a significant decrease of the PDs variance was found at the highest ND values, suggesting other sources of error than partial volume effects. It is concluded that with an ND value of about 3, accurate blood volume flow rates can be determined.

摘要

由于血管边界处的部分容积效应,预计在小血管中磁共振相位对比体积流量测量的准确性低于大血管。使用超声渡越时间流量计(TT)在直径为3.5±0.7毫米的犬股动脉中验证了准确性。每个血管直径的像素数(ND)范围为1.6至4.8。使用幅度图像中的阈值确定血管横截面。两种方法之间的相关系数为0.95(范围为10 - 200毫升/分钟)。比例差异(PD),即(QTT - QMR)/1/2(QTT + QMR),为0.8%,表明两种方法之间无系统差异。PD的标准偏差为27%,对于流速高于30毫升/分钟的情况为19%。仅在最高ND值时发现PD方差有显著下降,这表明除部分容积效应外还存在其他误差来源。得出结论,当ND值约为3时,可以确定准确的血容量流速。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验