The PFA-100 system (Dade International Inc., Miami, FL) is a platelet function analyzer, the design of which is based on the technology of the Thrombostat 4000 VDG, Seeon, Germany. It was developed to measure primary, platelet dependent hemostasis in citrated whole blood in vitro. A first pilot study was conducted with the instrument to assess performance characteristics. Healthy subjects (normals) who had not ingested any medications, and patients (abnormals) with primary, platelet-related hemostasis defects, which included users of aspirin, were studied with two test cartridges; collagen/ADP and collagen/epinephrine. Before the study certain variables were tested that ascertained that blood drawn into either 3.8% or 3.2% sodium citrate containing vacutainers (rather than syringes) could be used for testing. Tests must be performed within a five-hour time span from drawing to testing, and blood must be kept at room temperature. Normal reference values were 77-133 seconds closure times for collagen/ADP and 98-185 seconds for collagen/epinephrine. Precision testing revealed a CV of < 10% for within-day and between-day (five days) analyses on collagen/ADP cartridges and a CV of 5-14% for both runs on the collagen/epinephrine cartridges. No clinically important differences were found between measurements in the two positions of the instrument, although one follows the other.(ABSTRACT TRUNCATED AT 250 WORDS)