Suppr超能文献

对早晚气流的测量严重低估了哮喘患者第一秒用力呼气容积(FEV1)和呼气峰值流速的昼夜变异性。

Measurements of morning and evening airflow grossly underestimate the circadian variability of FEV1 and peak expiratory flow rate in asthma.

作者信息

D'Alonzo G E, Steinijans V W, Keller A

机构信息

Division of Pulmonary and Critical Care Medicine, Temple University School of Medicine, Philadelphia, PA 19140, USA.

出版信息

Am J Respir Crit Care Med. 1995 Sep;152(3):1097-9. doi: 10.1164/ajrccm.152.3.7663789.

Abstract

Peak expiratory flow rate (PEF) or forced expiratory volume in 1 s (FEV1) monitoring is an important clinical tool to assess the degree of 24-h variation (circadian) in lung function, which correlates with the degree of airway hyperresponsiveness. Ideally, PEF measurements should be taken frequently during the 24-h period to reflect the true variability. In practice, however, measurements are generally taken twice daily, usually upon arising and in the evening before bedtime. The effect of reducing the frequency of lung function measurement on the reported 24-h time average (mesor) and the circadian variability were evaluated. Four measurements taken 8 h apart and two measurements derived upon arising and the other, either 12 h later or before retiring, were compared with the full 24-h lung function profile, which included thirteen 2-hourly measurements. Based on the data from a 24-h lung function study in 20 adult patients with asthma, the corresponding values were separately compared at baseline and under two theophylline treatments, one being administered twice daily and the other once daily. The results are as follows: Whereas the 24-h time average (mesor) is hardly effected by the reduced measurements schemes, variability was grossly underestimated. Only 60 to 80% of the actual variability is picked up in the case of the four 8-hourly measurements, and only 20 to 45% in the case of the two measurements. The loss of accuracy in assessing 24-h airflow variability has potentially important implications on clinical practice and the investigation of therapeutics in asthma.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

呼气峰值流速(PEF)或一秒用力呼气容积(FEV1)监测是评估肺功能24小时变化(昼夜节律)程度的重要临床工具,该变化程度与气道高反应性相关。理想情况下,应在24小时内频繁进行PEF测量以反映真实变异性。然而,在实际操作中,测量通常每天进行两次,通常在起床时和晚上睡前。评估了减少肺功能测量频率对报告的24小时时间平均值(中值)和昼夜变异性的影响。将每隔8小时进行的4次测量以及起床时进行的2次测量(另一次在12小时后或睡前进行)与完整的24小时肺功能曲线进行比较,完整曲线包括每2小时进行一次的13次测量。基于20名成年哮喘患者的24小时肺功能研究数据,分别在基线以及两种茶碱治疗(一种每日给药两次,另一种每日给药一次)下比较相应值。结果如下:虽然减少测量方案对24小时时间平均值(中值)影响不大,但变异性被严重低估。对于每8小时进行一次的4次测量,仅能检测到实际变异性的60%至80%,而对于2次测量,仅能检测到20%至45%。评估24小时气流变异性时准确性的损失对哮喘的临床实践和治疗研究可能具有重要意义。(摘要截选至250字)

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验