Wang Y, Grimm R C, Rossman P J, Debbins J P, Riederer S J, Ehman R L
Mayo Clinic, Rochester, MN 55905, USA.
Magn Reson Med. 1995 Jul;34(1):11-6. doi: 10.1002/mrm.1910340104.
To reduce respiratory blur and ghosts in 3D coronary imaging, a data acquisition scheme using consistent multiple breath-holds was implemented. A navigator echo was acquired and processed in real time to dynamically measure diaphragm position. This information was provided as a visual prompt to the patient to maintain consistency in breath-hold levels such that the variation range of diastolic heart position was less than 2 mm. Preliminary results indicate that this multiple breath-hold acquisition scheme, compared with acquisition under respiration, can significantly reduce blur and ghost artifacts in 3D coronary imaging.