Determination of isotope ratios in human tissues enriched with zinc stable isotope tracers using inductively coupled plasma-mass spectrometry (ICP-MS).

作者信息

Friel J K, Longerich H P, Jackson S E

机构信息

Department of Biochemistry, Memorial University of Newfoundland, St. John's, Canada.

出版信息

Biol Trace Elem Res. 1993 May-Jun;37(2-3):123-36. doi: 10.1007/BF02783788.

Abstract

Enriched fecal and urine samples were prepared using ion-exchange column chromatography for analysis by Inductively Coupled and Fast Atom Bombardment Mass spectrometry (ICP-MS, FAB-MS) to compare precision between methods. Unenriched samples of human milk, feces, and whole blood were prepared similarly to monitor instrumental precision and analytical error. A least squares fit of the ICP-MS results vs the FAB-MS for 70Zn/64Zn gave a slope of 0.98, with a relative standard deviation (RSD) of only 0.7%. The results for 68Zn/64Zn gave a slope of 0.82, with a RSD of 14%. For unenriched tissues, all potential interferences were removed by the preparation procedure with no significant differences between preparation for isotope ratios of 70Zn/64Zn, 68Zn/64Zn, 67Zn/64Zn, and 66Zn/64Zn. Poisson counting statistics are a major contribution to the total analytical error indicating the usefulness of this procedure for enrichment studies.

摘要

使用离子交换柱色谱法制备富集的粪便和尿液样本,以便通过电感耦合和快原子轰击质谱法(ICP-MS、FAB-MS)进行分析,以比较不同方法之间的精密度。以类似方式制备人乳、粪便和全血的未富集样本,以监测仪器精密度和分析误差。ICP-MS结果与FAB-MS对(^{70}Zn/^{64}Zn)的结果进行最小二乘法拟合,得到斜率为0.98,相对标准偏差(RSD)仅为0.7%。(^{68}Zn/^{64}Zn)的结果斜率为0.82,RSD为14%。对于未富集的组织,通过制备程序消除了所有潜在干扰,(^{70}Zn/^{64}Zn)、(^{68}Zn/^{64}Zn)、(^{67}Zn/^{64}Zn)和(^{66}Zn/^{64}Zn)同位素比的制备之间无显著差异。泊松计数统计是总分析误差的主要贡献因素,表明该程序在富集研究中的有用性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验