van't Riet A, Mak A C, te Loo H J, Hoekstra C J, Althof V G
RISO, Deventer, The Netherlands.
Radiother Oncol. 1994 Nov;33(2):167-70. doi: 10.1016/0167-8140(94)90071-x.
In brachytherapy afterloading, the treatment is simulated using dummy sources. In this study, imaging during treatment was performed to see whether the radioactive sources accurately replaced the dummy sources. In low-dose rate afterloading (manually or remote controlled) X-ray radiography was applied for verification. In high dose rate afterloading, the radioactive source was visualised under remote controlled X-ray fluoroscopy. Differences between dummy and treatment source position were observed, justifying verification imaging in brachytherapy.